Rerun Key; Print Key; Export Key - SA Instruments 1030 Operation Manual

Mr-compatible small animal monitoring and gating system
Table of Contents

Advertisement

SnapView

ReRun key

Click the ReRun key on the SnapView display to play back the 36 second data set.

Print key

The print key will print a frozen display. The desired printer should be selected as the
default printer and the paper orientation should be set to landscape before running
SnapView.

Export key

The Export key allow SnapView's raw data to be exported. This feature can be useful in
the event wave shapes or features need to be analyzed offline.
Keying the EXPORT key opens an export setup window with preset parameters. The
user can modify the parameters and make measurement selections before exporting
the data to a .txt file. Up to 32766 lines of 19 items comma delimited can be exported to
a .txt file. Each text line starts with a time sample number (0-32766) followed by a raw
data count for the measurements selected.
If SnapView is at the end of displaying the data set and EXPORT is keyed, then the
export routine is setup to dump the entire data set. The resulting file when opened and
saved by Excel will be about 4.6 Mbytes.
If SnapView is running (or FROZEN) and EXPORT is keyed the export parameters are
setup to output data from the beginning of the data set up to the current (or frozen) time
point. Note the time counter SS:HH (seconds: hundredths of seconds) displays the time
of the data set from 0 to 36.4 seconds.
Revision C, August 2013