Appendix D Routine Maintenance - SA Instruments 1030 Operation Manual

Mr-compatible small animal monitoring and gating system
Table of Contents

Advertisement

Appendix D
Routine Maintenance
Clean the monitor's surfaces or probes with a soft cotton cloth moistened in a mild soap
solution. If disinfection is required, wipe the surface with alcohol, sodium hypochloride
or glutaraldehyde (Cidex, Metrucude, etc.). Follow the instructions provided by the
manufacturer.
Charge the ERT Battery Pack when the system is not in use. The battery pack can be
left attached to the charger after it is fully charged. There is no requirement to deep
discharge the battery before charging.
Charge the battery in the IBP Module when the system is not in use. The charger is
internal to the module. The internal charger automatically charges the internal battery
when the IBP Module is connected to 12 VDC power.
The water, water filter and roller pump tubing in the Water Bed Heater System should
be changed annually or as needed. Refer to the pictures below of the Circulation
Module with the covers removed.
When removing tubes from the Heater Module, clamps can be useful on the 3/8" OD
tubing. Plugs can also be used on the Heater Module ports.