Fiber Optic Temperature

PC-sam's main display

When the "all temperatures" selection is made in the MONITOR SETUP window, a row
of buttons appears below the temperature trend in PC-sam's main display. These
buttons allow the user to select which temperatures are displayed in the trend and
which current temperature is displayed in the digital display to the right of the trend.
In the example above, both T1 and FOT1 have been selected and are present in the
trend display. FOT1 was selected last, so that probe's current measurement is also
displayed. The current measurement can be switched to T1 by left clicking on the T1
button.
To remove a measurement set from the trend display right click on the
corresponding button.

Heater control

The SA Instruments warm air Heater System can be controlled by the Fiber Optic
Temperature module. The HEATER window can be used to select which temperature
probe will be used to control the heater and to set the set point. Refer to Chapter 6 for a
detailed explanation of the Heater System.
For the module with LEDs and buttons, the digital display on the Fiber Optic
Temperature Module can display the current temperature, the set point or the PWM
duty factor. Two buttons on the face of the module facilitate which value is displayed.
Revision C, August 2013