Advanced Features
Note that the Apply and Cancel keys control the values entered for the auto track
parameters. Pressing the Cancel key will display the previous value while pressing
Apply sends the value to the ERT Control/Gating Module. The background color in the
value box indicates the status. White indicating the value has been sent to the ERT
Control/Gating Module. If the enabled box is not checked it does not matter if the
parameter value has been sent or not as it will not be used to determine the gate
position or width.
For the example shown above Auto Track has been selected to apply to Respiration.
The respiration gate has been defined to be during the expiration portion of the
waveform by checking Pulse Invert. Begin Delay and Max Width have been set to 5
and 50% of the respiration interval or 15.5 and 155 msec respectively. When changes
occur in the animal's respiration rate, the two parameters will automatically track
according to the parameter settings.
Revision C, August 2013
15-7