Overview; Fiber Optic Communication Cable Connections; Ibp Transducers - SA Instruments 1030 Operation Manual

Mr-compatible small animal monitoring and gating system
Table of Contents

Advertisement

Overview

The IBP Module measures invasive blood pressure using a blood pressure transducer
connected with tubing to an invasive line in the animal. System design accommodates
up to three IBP channels with one channel designated as IBP3 incorporated in the IBP
Module. The other two IBP channels (IBP1 & IBP2) are either incorporated into the IBP
Module or they are battery-powered modules which daisy chain with simplex fiber optic
cables from IBP3. The IBP Module has two auxiliary inputs to allow the user to input
event markers or acquire, record, display and gate from user generated analog
waveforms.
The module has bi-directional fiber optic communication to the ERT Control/Gating
Module and is powered from a 12 VDC external supply or by internal battery.
Caution: the 12 VDC power supply is slightly magnetic and
should be fix mounted in the fringe field of the
magnet.
IBP is an optional feature on the Model 1030.

Fiber optic communication cable connections

When the IBP Module is the only optional module, it connects to the ERT Control/Gating
Module using a duplex fiber optic cable. The cable should be connected to the fiber
optic ports labeled "option in" and "option out" and the options I/O switch should be set
to "options in use".
connect to form a data loop with one of the duplex fibers connecting to the 1
module and the other of the duplex fibers connecting to the last optional module. The
modules are then connected to each other using short simplex fiber optic cables.
The fiber optic cables have blue and grey connectors. The driver and receiver ports are
also blue and gray. The order of the modules in the data loop is not important. But, in
all cases the color of the connector must match the color of the port
.

IBP transducers

The Model 1030 is configured to operate using disposable invasive blood pressure
transducers which meet the specification of 5µV/V/mmHg.
transducers meet this specification and can be used with the Model 1030.
When other optional modules are also present, the modules
Revision C, August 2013
Invasive Blood Pressure
The following IBP
Chapter 10