ALARMS window

The ALARMS window allows high and low alarm limits to be set for heart rate, oxygen
saturation, respiration rate and/or temperature. A check mark next to an alarm limit
followed by clicking the Apply key will activate the alarm. When the measured value
meets or exceeds the limit, an audible alarm sounds.
The window has an area for messages and keys for Quick Set Alarms and Alarm
Silence. The Quick Set key provides a convenient way to quickly set all alarm limits.
Clicking the Quick Set key sets the high and low alarm limits 5% above and below the
current measurement for heart rate, oxygen saturation and respiration rate and 2 ºF
above and below the current temperature measurement.