Water Bed Heater System

Water Beds

There are several options for water beds. The simplest is just tubing lying beside or
wrapped around the animal. Many animal handling systems have tubes for fluids that
can be used as a water bed.
Water beds available from SA Instruments include flat and shell beds for mice and rats.
Shell beds have the shape of an inverted U and can cover the animal. The beds are
inserted in a thin disposable plastic cover before use.
Shown below are a shell beds for mice and rats and a flat bed for a rat.

Water Bed Heater System setup in MR

Refer to Appendix D for maintenance instructions for the Water Bed Heater system.
The Circulation Module (8801000 should be located outside the RF shielded magnet
room. The Circulator Module has magnetic parts and is not RF shielded. The Water
Bed Heater Module (76x100) should be located inside the RF room near the bore of the
magnet. The Heater Module is RF shielded and non-magnetic. It is powered from 100,
115 or 230 VAC, 50 – 60 Hz. Check to make sure your power matches the specification
on the Heater Module label.