Driving
1. Give drive command in desired direction (forward or
reverse).
2. Release drive command.
Wheelchair speed accelerates to 20% of the maximum
drive speed.
3. Give forward command when driving forward or reverse
command when driving in reverse to accelerate to next
fixed speed.
4. Release drive command.
New speed is held constantly.
Stopping
Use one of the following methods to stop:
Give drive command in opposite direction (a reverse
command when driving forward or a forward command
when driving in reverse)
Press external stop switch
Perform emergency stop
Let Latch Drive Timeout expire
60101834-C

6.6.5 3 Step Up/Down

In this mode, you can step up or down through one of three
fixed speeds. The speeds available are 33%, 67% and 100%
of the maximum pre-set reverse or forward speed A of the
selected drive screen and then remain at that speed for
the programmed Latch Drive Timeout period as long as no
further command is given.
Driving
1. Give drive command in desired direction (forward or
reverse).
2. Release drive command.
Wheelchair speed accelerates to 33% of the maximum
drive speed.