Combination operation
3.1 Analyzer module/slot combinations
Note
Influence of building vibrations on the measured signal
If you are using OXYMAT 7 and/or ULTRAMAT 7 analyzer modules, building vibrations, in
particular in the frequency range from 6 to 15 Hz, can influence the measured signal.
• Select an installation location which is vibration-free.
• If necessary, change the measuring frequency settings:
– Make sure that the measuring frequencies do not correspond to the building vibration
– Also make sure that the measuring frequency is not the same as half the fundamental
3.1.2

Analyzer module/slot combinations: Slots

Slots for analyzer modules
The figure below shows the slots for the analyzer modules:
R
W
Figure 3-1
16
or to one of its harmonic multiples.
frequency of the building vibration.
Rack-mounted device (rear)
Wall-mounted device (bottom)
Slot 1 (AM1)
Slot 2 (AM2)
Slots for analyzer modules
Compact Operating Instructions, 06/2017, A5E31805656-05