Table 3- 6
Case
PM
New
New
Data synchronization: PM parameters are transferred to the AMx (x=1 or 2), parameters of the
1)
AMx are transferred to the PM
Serial number stored in the PM is also the serial number of the device
2)
Table 3- 7
Case
PM
Re-
placed
Re-
placed
Data synchronization: PM parameters are transferred to the AMx (x=1 or 2), parameters of the
1)
AMx are transferred to the PM
Quick Start
Compact Operating Instructions, 06/2017, A5E31805656-05
Plug & Measure installation of a PM into an existing (configured) device / replacement of
a PM in an existing (configured) device
Description
AMx
Un-
The PM electronics is replaced. "-" is
changed
displayed instead of the serial number.
At least one AM remains in the device.
Un-
PM electronics with new serial number
changed
is installed.
Plug & Measure - Replacement of a PM/Installation of a used AM
Description
AMx
Re-
Assembly from inventory: Each module
placed
has a serial number and has been
connected to another module before.
AM of the same type is replaced.
Re-
Assembly from inventory: Each module
placed
has a serial number and has been
connected to another module before.
AM of a different type is replaced.
Combination operation
3.4 Plug & Measure: Permissible activities
Data synchronization/Action
Automatic data synchronization
action required by user.
When there are two AMs in the device,
the new PM always synchronizes with
AM1.
The PM receives the data of the previ-
ously used PM, including its serial
number
2)
Automatic data synchronization
action required by user.
In this scenario, the replacement of the
PM electronics results in a loss of data
of the previously installed PM. In addi-
tion, the serial number is overwritten.
Data synchronization/Action
User action. Confirm with <ENTER> or
reject with <ESC>.
When the process is confirmed, stored
AM data is transmitted to the replaced
AM.
Automatic data synchronization
action required by user. Each module
retains its data.
, no
1)
, no
1)
, no
1)