Commissioning: Sequence
8.2 Procedure
1. Supply the device with power. Information on the power supply of the device is available
in the device-specific documentation → References (Page 108).
The boot process starts. The splash screen is initially visible on the device display.
2. Scan of basic settings:
The boot process is interrupted momentarily. Two successively displayed editing windows
prompt you to select the display language and to enter the date and time.
– Select one of the offered language options in the first editing window.
– Then set the date and time of day.
Once you have entered the requested information, the boot process continues.
3. Wait for transition to warming-up phase:
When the splash screen disappears, the device is in the warming-up phase. The current
temperature status is displayed as a bar graph in the warming-up window.
4. Call the measured-value display during the warming-up phase
– Close the warming-up window by pressing <ENTER>.
88
Note
Date / time settings
When the power is off, these settings for the date and time are stored on the device
for a period of approximately 14 days.
After putting the device back into operation, you will need to reset the date and time.
The display shows the measured-value display in read mode.
For the duration of the warming-up phase, the device is in "Function check" mode.
The following symbol is displayed in the status bar:
Compact Operating Instructions, 06/2017, A5E31805656-05
Quick Start