Note
Authorization level
To execute the actions described below, you must be logged on with at least the "Expert"
authorization level. You may have to log on again with the requested PIN.
Note
Alarm window
When you change the hardware of the device by replacing or adding modules or
components, the alarm window "Device configuration changed" is displayed in the LUI during
re-commissioning. To start the data synchronization, confirm the alarm by pressing
<ENTER>.
Depending on the specific application, parameter sets may have to be transmitted. The
message window "Transmit parameters" is displayed. Information on the further procedure is
available in the tables below:
→ Case differentiations (Page 30)
→ Case differentiations (Page 30)
→ Case differentiations (Page 30)
→ Case differentiations (Page 30)
→ Case differentiations (Page 30)
Quick Start
Compact Operating Instructions, 06/2017, A5E31805656-05
Combination operation
3.4 Plug & Measure: Permissible activities