Combination operation
3.3 Gas connections
Retrofitting example: "Pressure operation, serial" for OXYMAT 7/ULTRAMAT 7
When you are replacing the ULTRAMAT 7 with an OXYMAT 7 analyzer module, you must
replace the sample gas restrictor in the sample gas inlet with a clamping screw.
MGE, VGE
①
②
③
④
Figure 3-5
26
Sample gas inlet (MGE), Reference gas inlet (VGE)
Initial situation, serial pressure operation of OXYMAT 7 (AM1) and ULTRAMAT 7
(AM2)
Remove sample gas restrictor in the sample gas inlet of the replacement module
Install clamping screw in the sample gas inlet of the replacement module
Replace ULTRAMAT 7 with OXYMAT 7
Retrofitting of the sample gas restrictors: Example "Pressure operation, serial"
Compact Operating Instructions, 06/2017, A5E31805656-05
Quick Start