Setting up
Note
Operator control options
When commissioning for the first time, two methods can be used. You can commission the
device with the simplified process using the "[1] Quick Start" menu or opt for the more
detailed procedure described below.
• Additional information on the operating steps of the detailed process is available in the list
of references →Table A-1 References 1 - Operating Manuals LUI (Page 108).
• When you are working with the "[1] Quick Start" menu, you set the parameters of the
following menus one after the other:
→[1.1] Basic settings (Page 91)
→[1.2] Automatic log-off (Page 92)
→[1.3] Analog outputs (Page 93)
→[1.4] Measuring ranges (Page 95)
→[1.5] Calibration (Page 99)
→[1.6] Save parameter set (Page 105)
Short description of the detailed process
If you do not use the quick start procedure, commission the device with the steps described
below:
1. Call the main menu:
– Switch the measured-value display to selection mode. To do so, press <ENTER>.
– Press <Enter> again to open the main menu.
2. Identify device: → [3.09] Maintenance & Diagnostics > Identification.
3. Specify the PIN for the user levels:→ [5.1] Security > Access management.
In the factory settings, the PIN 1111 is the default for the "Standard" user level and PIN
2222 for the "Expert" user level.
4. Disable automatic logoff temporarily: The factory setting is a logoff time of 5 min. When
this time elapses, changes that were not stored retentively are lost. We recommend that
you temporarily deactivate the automatic logoff during commissioning.
For additional information, refer to: → [5.2] Security > Automatic logoff.
Quick Start
Compact Operating Instructions, 06/2017, A5E31805656-05
Commissioning: Sequence
8.2 Procedure