Siemens SIPROCESS GA700 Quick Start Manual page 24

Continuous gas analysis
Hide thumbs Also See for SIPROCESS GA700:
Table of Contents

Advertisement

Combination operation
3.3 Gas connections
OXYMAT 7
Note
Order additionally required clamping screws separately
Additionally required clamping screws are not included in the scope of delivery. They must
be ordered separately.
Table 3- 1
Operating mode
Pressure operation
Suction operation
24
Mounting conditions for the sample gas restrictor: OXYMAT 7
Sample gas restrictor/clamping screw
Operation with one
OXYMAT 7
In front of the sample
gas inlet
In the sample gas outlet In the sample gas outlet
Operation with two
OXYMAT 7 in series
In front of the sample
gas inlet into which the
sample gas is intro-
duced first:
Replace the sample
gas restrictor in the
inlet of the rear, se-
cond module with a
clamping screw
from which the sample
gas is first sucked out:
Replace the sample
gas restrictor in the
inlet of the rear, se-
cond module with a
clamping screw
Compact Operating Instructions, 06/2017, A5E31805656-05
Operation with two
OXYMAT 7 in parallel
In front of each sample
gas inlet
At each sample gas
outlet
Quick Start