Figure 3-9 Analyzing Patient Size Display Panel Screen - ZOLL AutoPulse 100 User Manual

Resuscitation system
Table of Contents

Advertisement

2.
Press and release the Start/Continue button once. The Platform automatically adjusts the bands to
the patient's chest (see Figure 3-9).
Warning: Do not touch the patient or the LifeBand while the Platform is analyzing the patient's size.
Figure 3-9
Analyzing Patient Size Display Panel Screen
3.
The Platform will pause for 3 seconds to allow you to verify that the patient is properly aligned and
that the LifeBand has taken up any slack in the bands (see Figure 3-10).
If the patient is not properly aligned, press the Stop/Cancel button, realign the patient, and begin
compressions again with step 1 on page 3-7.
Warning: If the Stop/Cancel button is not pressed within 3 seconds, compressions will automatically
begin. Press the Stop/Cancel button to immediately stop the compressions.
Page 3-8
3 0 : 2
A n a l y z i n g P a t i e n t s i z e . . .
D O N O T T O U C H P A T I E N T .
STOP
User Guide