Table 2-2. R Series Controls and Indicators (continued)
Control or Indicator
4:1 Button
(optional)
NIBP Button
(optional)
Softkeys
Charge Indicator Light
(not shown)

Display Screen

The front panel includes a color display which shows:
The elapsed time (since the unit was turned on).
The ECG trace, selected lead, size, heartbeat indicator, and alarm status.
The selected energy, charging status, and delivered energy for defibrillation and
synchronized cardioversion.
The output current and stimulus rate for pacing.
The measured SpO
applicable), and alarm status indicators for optional SpO
Non-invasive blood pressure (NIBP) readings: diastolic, systolic, and mean, plus alarm
status indicators (optional; refer to the insert Non-Invasive Blood Pressure (NIBP), part
number 9650-0914-01).
The patient's carbon dioxide level, respiration rate and capnogram (if applicable), and alarm
status indicators for CO
(EtCO
Messages and prompts.
Labels above the softkeys (appropriate to the context).
CPR Index™ and Release Bar.
CPR Rate and Depth.
Figure 2-2 shows the layout of parameter values, waveforms, system data, and softkey labels.
9650-0912-06 Rev. G
Description
This button is used to determine a patient's underlying ECG rhythm.
While depressed, this button causes pacing stimuli to be delivered at ¼
of the indicated ppm setting. When the button is released, normal
pacing resumes.
Allows you to start single, auto, or STAT non-invasive blood pressure
measurements as described in the option insert Non-Invasive Blood
Pressure (part number 9650-0914-01). Your unit has this button only if
you ordered this configuration.
Six unlabeled buttons located directly below the display control different
functions depending on the operating mode of the unit.
Labels for the softkeys appear at the bottom of the display directly
above each softkey to indicate its function.
Located on the apex paddle, this light turns on when the defibrillator is
charged and ready.
percent saturation, signal strength, plethsymographic trace (if
2
monitoring (optional; refer to the insert End Tidal Carbon Dioxide
2
), part number 9650-0915-01).
2
ZOLL R Series Operator's Guide
Defibrillator Controls and Indicators
monitoring.
2
2–5