Warning:
If a user advisory or fault cannot be cleared or a system error occurs during active operation,
immediately revert to manual CPR.
Operating the Platform on a patient for extended periods of time may result in minor skin
irritation to the patient.
Do not use the Platform in the presence of an oxygen-rich (greater than 25% oxygen)
atmosphere, flammable anesthetics, or other flammable agents (such as gasoline). Using the
Platform near the site of a gasoline spill may cause an explosion.
To avoid the risk of electric shock, connect the AutoPulse Multi-Chemistry Battery Charger
(Battery Charger) only to a supply mains with protective earth.
Always charge a new Battery upon receipt. Failure to charge a Battery may cause reduced
Battery performance.
Always charge a stored Battery before placing the Battery into active operation. The Battery
slowly self-discharges when not in use. Failure to charge a Battery before use may cause
device power failure. In no case should any Battery be used if it has not been charged within
the previous 60 days.
No modification of the AutoPulse Platform, the LifeBand, or the AutoPulse Power System
is allowed.
Caution: United States federal law restricts this device to sale by or on the order of a licensed physician.
Caution: The AutoPulse System is designed to be used only with ZOLL-approved accessories. The
AutoPulse System will perform improperly if non-approved accessories are used.
Caution: Only use ZOLL Batteries specifically designed for use with the AutoPulse System. The use
of other batteries may cause permanent damage to the AutoPulse System and will void the
warranty.
Caution: Do not store an AutoPulse Battery in an unpowered Battery Charger (unplugged or AC power
is off). Irreversible damage to the Batteries occurs within 10 days, depending on the initial
state of charge.
Caution: Do not autoclave the Platform, the LifeBand, or the AutoPulse Power System.
Caution: Retain the original product literature for future reference.
Caution: Do not use or stack the unit with other equipment. If the unit is used or stacked with other
electrical equipment, verify proper operation before using it.
Caution: Any serious incident that has occurred in relation to the device should be reported to the
manufacturer and the competent authority of the Member State in which the user and/or
patient is established.