User Guide
Figure 3-8
Fastening the LifeBand
3.2
Starting Chest Compressions
1.
Make sure that the yellow upper edge of the LifeBand is aligned with the patient's armpits, and is
directly over the yellow line on the Platform. Also make sure that there are no obstructions, such as
clothing, straps or equipment, with the bands.
Warning:
•
Failure to properly position the LifeBand at the patient's armpit line may cause injury to the
patient.
•
Failure to properly position a patient, both vertically and laterally with respect to the
Platform, may cause injury to the patient.
•
Do not strap across, or otherwise constrain, the LifeBand. Constraining the movement of the
bands can damage or break the LifeBand.
Page 3-7