ZOLL AutoPulse 100 User Manual page 64

Resuscitation system
Table of Contents

Advertisement

6.
Perform battery rotation after every use, or at least once every 24 hours. Use either the Three-
Battery Rotation or the Four-Battery Rotation method, as follows:
Three-Battery Rotation: Place the Battery that was removed from the Platform in step 2 into
the Battery Charger for charging. Check the spare Battery for Green LEDs, and then place it
into the Platform. Take a fully charged Battery from the Battery Charger, and use it as a spare
Battery.
Four-Battery Rotation: Place the Battery that was removed from the Platform in step 2 into the
Battery Charger for charging. Remove the spare Battery, and place it into the Battery Charger.
Take two fully charged Batteries from the Battery Charger, check for Green LEDs on each
Battery, and then place one into the Platform, and use the second as a spare.
7.
Perform AutoPulse Self-Test daily:
1.
Ensure that a LifeBand is installed.
2.
Ensure that a fully charged Battery is installed, and power on the Platform.
3.
On power up, all LEDs on the display will momentarily light up, and then only the Green
POWER LED will remain lit.
4.
The Battery Charge Status Icon on the Platform User Control Panel should also be visible with
4 bars; if not, replace the Battery with a fully charged Battery from the Battery Charger.
5.
If the Platform RED ALERT LED remains on, refer to Chapter 5, "Troubleshooting
Procedures". If the RED LED can not be resolved, contact ZOLL.
Inspect the Battery in accordance with the inspection checklist provided below. Follow the
recommended actions as appropriate. The Battery is automatically checked each time the device is
powered on.
1.
Inspect the Battery, including the connector for physical damage that would preclude its insertion
in either the platform or the Battery Charger. Contact ZOLL as needed.
2.
Inspect the Battery for cracks in the Battery case exposing internal components. If the Battery is
damaged, do not attempt to place the Battery into the Platform- this can cause damage to the internal
connector of the Platform. If damaged, do not use.
Inspect the Battery Charger in accordance with the inspection checklist provided below. Follow the
recommended actions as appropriate.
1.
Inspect the Battery Charger for physical damage. Contact ZOLL as needed.
2.
Inspect the battery compartments for foreign substances, be sure they are free and clear for battery
insertion. Remove any foreign matter or debris.