Branson Sonifier SLPt User Manual page 51

Cell disruptor
Table of Contents

Advertisement

SLPt Cell Disruptor
User Manual
Table 7.1
System Trouble Analysis Chart
Symptom
Ultrasonic power is
absent or inconsistent
Sonifier operates continu-
ously but does not oper-
ate in Pulsed mode
Slight electric shock when
touching a metal part of
the system or lab equip-
ment contacting the sys-
tem
User I/O signals are not
working correctly
100-214-252 Rev. A
DOC EXPIRES 12PM 9/16/2007. Article or Material must comply with the requirements
stipulated by RoHS in its current version of 1/27/03
Probable Cause
• Tool is loose or has failed
• Converter cable connection
is loose or has failed
• Converter has failed
• Power supply or controls
have failed
• Digital controls have failed
• System is not properly
grounded
• Cordset has failed or had
Ground lead removed
• User I/O is not configured
correctly
• customer-provided User I/O
components have failed or
are no longer functioning
• Outputs of User I/O failed
Chapter 7: Maintenance
Troubleshooting Charts
Corrective Action
• Tighten or replace
defective tool
• Tighten converter cable
on back of Sonifier
• Replace defective Con-
verter, return for repair.
• Return for repair
• Return for repair
• Correct electrical ground
to system
• replace cordset
• Verify and correct con-
nections - See Section
Table 4.2, User I/O Pin-
Out (DB9F) provided
for customer-designed
interface, beginning on
page 6-9
• Return unit for repair
7-5