Chapter 2: Introduction to the SLPt Cell Disruptor
Controls and Commands

2.2.1 SLPt Cell Disruptor Keypad/Display Description

LED Display
A 4 digit LED numeric display is located on the front panel. It is controlled by the Left and Up arrow keys
located adjacently on the membrane keypad, and are used to respectively select and increment the selected
digit. This allows setup and navigation for controlling the SLPt.
Mode Select Key
Each of the 4 SLPt Modes (Continuous, Time min/sec, Time hr/min, Pulse) can be selected by pressing the
Mode key. Repeat the press to advance to the desired operational mode. The selected Mode LED will light.
Set Key
Use this key to select the parameter(s) you want to enter/update on the display (LED(s) 3A, C, D, E).
Reset Key
When an alarm occurs during the process cycle, and the Reset LED is blinking, pressing the Reset key will
clear the alarm. If the LED is not on, and RDY is displayed, pressing the Reset key will send a reset to the
power supply module.
When in the parameter entry mode, pressing the Reset will restore the original value of the parameter being
modified and then return the system to Ready. It will also reset the power supply and clear any alarms.
Test Key
The Test key is only active when the system is ready. Ultrasonics are enabled while the Test key is pressed,
and disabled when released. Test performs the run, seek, and store functions, causing the power module to
test for the resonant frequency of the converter and tool at 10% amplitude, and then go to the amplitude set-
ting.
Enter Key
Press the Enter key on the SLPt Cell Disruptor to:
Confirm a selection setting.
Switch between power display and experiment time remaining
Exit the configuration menu.
If you have entered a modified parameter value, the system checks the value for validity and format. If the
value you entered is valid, the value is accepted, and it replaces the old value. The original value is lost and
cannot be recovered. If the modified value is invalid, an appropriate error message is displayed.
Start/Stop Key
The Start/Stop key is used to set up Pulse Start in the configuration menu at start up. Either: Press and hold
to run a cycle for its duration, or Press and release to start a cycle . . . press again to abort cycle.