User Manual

4.6.2 User I/O Connection

The SLPt Cell Disruptor is equipped with a standard external connection to allow you to design and connect
your own custom interface for controlling the unit. The User I/O interface can be useful when you need to
activate the Sonifier remotely, for example, when the operator must start and stop the unit from another
room for safety reasons.
Table 4.2
User I/O Pin-Out (DB9F) provided for customer-designed interface
DB9F Pin
1
2
3
4
5
6
7
8
9

4.7 Guards and Safety Equipment

Although the SLPt Cell Disruptor operates outside the normal range of human hearing, some applications
can create audible noise above 85dB. If an uncomfortable level of noise is present, the operator should wear
ear protection for safe operation.
Appropriate eye protection should be worn when operating the SLPt, to prevent possible splash injury orig-
inating in the solution.
WARNING
!
The Ultrasonic Horn can cause injury and/or equipment damage during operation. To avoid
injury or accident, never touch the Ultrasonic Horn while the System is turned on, and do not
allow the Horn to come in contact with solid vessels or supports.
The User I/O may be used to remotely control the System. If this is the case, you must design in whatever
safety precautions are appropriate to your User I/O circuit design to prevent unexpected start-up, which can
cause personal injury and can cause equipment damage.
100-214-252 Rev. A
DOC EXPIRES 12PM 9/16/2007. Article or Material must comply with the requirements
stipulated by RoHS in its current version of 1/27/03
Function
Reset
Start
Weld On
Alarm
Ready
Start Source
Reset, Alarm, Ready, Weld On
Seek
Seek
Chapter 4: Installation and Setup

Guards and Safety Equipment

Signal
Input
Input
Output
Output
Output
Input
Common
+24 Volts Input
Ground
4-9