Slpt In Time Mode (Either Low, Min/Sec Or High, Hr/Min); Slpt In Time Mode With Pulse (Either Low, Min/Sec Or High, Hr/Min) - Branson Sonifier SLPt User Manual

Cell disruptor
Table of Contents

Advertisement

Chapter 6: Operation
Operational Sequence

6.4.2 SLPt in Time Mode (Either Low, min/sec or High, hr/min)

1. Turn On power, wait for display (1) to show Ready (rdY).
2. Select Mode (2) > Time (2B). The On (3A), Off (3C), and Amplitude LED's (3D) will light.
3. Set the On time, Off time and Amplitude % using the arrow keys.
4. Press ENTER.
5. Send Start pulse by shorting pins 2 and 6 of external user I/O, J2 to start cycle, or by pressing the Start/
Stop key, as per configuration.
6.4.3 SLPt in Time Mode with Pulse (Either Low, min/sec or High, hr/
min)
1
88:88
2
3
Reset
4
WARNING
!
Do NOT touch the vibrating horn, or place the vibrating horn against solid objects. Touching
or holding the horn can result in burns or injury, and contacting solid materials with the vibrat-
ing horn or tip can cause breakage.
6-8
DOC EXPIRES 12PM 9/16/2007. Article or Material must comply with the requirements
stipulated by RoHS in its current version of 1/27/03
Continuous
Time(min/sec)
Mode
Time(hr/min)
Pulse
On
Set
Off(sec)
Amplitude(%)
ExpTime(hr/min)
Autotune
Start/Stop
DO NOT REMOVE
Test
Enter
5
6
7
9
1. Turn On power, wait for display (1) to show
Ready (rdY).
8
2. Select Mode (2) and press until Pulse (2E) and
Time (2B) are both selected. The On (3A), Off (3C),
Amplitude (3D), and Experiment Time (3E) LED's
will light.
2A
2B
3. Set the Time On, Off time, Amplitude %, and Exp
2C
time using the arrow keys.
2E
4. Press ENTER.
3A
5. Send Start pulse by shorting pins 2 and 6 of exter-
3C
3D
nal user I/O, J2 to start cycle, or by pressing the Start/
3E
Stop key, as per configuration.
100-214-252 Rev. A