Chapter 4: Installation and Setup
System Component Description
The tapered microtip attaches directly to the converter. The tapered tip is recommended for applications
such as spores, fungi, yeast, muscle, and connective tissue. Excellent results can be obtained on volumes
ranging from 3 to 20 ml in a relatively short period of time. As the diameter of the tip decreases, the ampli-
tude increases often requiring lower power settings for small volumes.
To prevent foaming or aerosoling while processing small quantities with the tapered microtips, the use of a
conical-shaped tube such as a reaction vial or a cut down centrifuge tube is recommended.
CAUTION
!
When using microtips, do not exceed a maximum Amplitude Control setting of 70%. The
microtip will break if driven at higher amplitudes.
CAUTION
!
The bottom of the test tubes should not be in contact with the surface of the ultrasonic tool.
Such contact could cause breakage and loss of sample.
4-4
DOC EXPIRES 12PM 9/16/2007. Article or Material must comply with the requirements
stipulated by RoHS in its current version of 1/27/03
100-214-252 Rev. A