3.2.2.1
Response CV
Slow: This option reduces the voltage output response speed with higher priority on
output stability.
Fast: This option increases the voltage output response speed with higher priority on
output response speed (default at power on).
3.2.2.2
Response CC
Slow: This option reduces the current output response speed with higher priority on
output stability.
Fast: This option increases the current output response speed with higher priority on
output response speed (default at power on).
3.2.2.3
V Limit
Select the V Limit Max/Min corresponding space to set the value.
Use this option to narrow down its range by setting the MIN and MAX. When setting the
output voltage, the DC power supply allows setting the voltage within the range of [MIN value
≤ user-defined value ≤ MAX value] which is V LIMIT: MAX=100V, MIN=20V. If the setting
exceeds the output voltage 110V set by the user, the BUZZER will beep one time (if
BUZZER is set to ON) and the main screen will automatically prompt a warning message.
3.2.2.4
I Limit
Select the I Limit Max/Min corresponding space to set the value.
Use this option can narrow down its range by setting the MIN and MAX. When setting the
output current, the DC power supply allows setting the current within the range of [MIN value
≤ user-defined value ≤ MAX value] which is I LIMIT: MAX=20A, MIN=2A. If the setting
exceeds the output current 21A set by the user, the BUZZER will beep one time (if BUZZER
is set to ON) and the main screen will automatically prompt a warning message.
3.2.2.5
P Limit
Select the P Limit Max/Min corresponding space to set the value.
Use this option can narrow down its range by setting the MIN and MAX. When setting the
output power, the DC power supply allows setting the current within the range of [MIN value
≤ user-defined value ≤ MAX value] which is P LIMIT: MAX=20W, MIN=2W. If the setting
exceeds the output current 21W set by the user, the BUZZER will beep one time (if BUZZER
is set to ON) and the main screen will automatically prompt a warning message.
Manual Operation
3-7