3.2.4.4.5 APG Current Output Calibration

Table 3-9 lists the hardware requirements for APG current output calibration.
Device
DVM
DC Power Supply
Figure 3-73 shows the wire connection for APG voltage output calibration.
When conducting the APG current output calibration, each calibration
point has to enter at least 4 significant digits to ensure the power supply
accuracy after calibration.
Calibration Procedure (Model 62180D-600):
Tap "Menu", "Configuration" and "Interface" to select "APG" page, set the I SOURCE
1.
SET and I LOAD SET to Verf (0-10V).
In Calibration page, tap "Host V" to select APG I SET and list the APG current
2.
calibration items as shown in Figure 3-77.
3-50
Table 3-9
Suggest Model or Capacity
HP 34401A or equivalent DVM
Any DC power supply or DC signal source that
can output 10Vdc and drive more than 100mA.
Figure 3-76