General Error Messages; Timeout - Agilent Technologies Infinity 1260 User Manual

Refractive index detector
Hide thumbs Also See for Infinity 1260:
Table of Contents

Advertisement

7
Error Information

General Error Messages

General Error Messages

General error messages are generic to all Agilent series HPLC modules and
may show up on other modules as well.

Timeout

The timeout threshold was exceeded.
Probable cause
1
2
100
The analysis was completed successfully,
and the timeout function switched off the
module as requested.
A not-ready condition was present during a
sequence or multiple-injection run for a
period longer than the timeout threshold.
Suggested actions
Check the logbook for the occurrence and
source of a not-ready condition. Restart the
analysis where required.
Check the logbook for the occurrence and
source of a not-ready condition. Restart the
analysis where required.
Agilent 1260 Infinity RID User Manual

Advertisement

Table of Contents
loading

Table of Contents