Switching Motion Orbital To Reciprocal In Sk-600 - Jeio tech Lab Companion SK-300 Operating Manual

Shaker
Table of Contents

Advertisement

Switching motion orbital to reciprocal in SK-600

-
Lower Spacer
Install position
1. Remove the Shaking table from the unit.
Remove M5x10, wrench bolt (4mm, wrench handle) from the Shaking table.
2. Remove M5*28L, Spacer (13mm, spanner) from the upper vibrating frame.
The upper vibrating frame is designated to move as the orbital motion.
3. Install M5*43L, Spacer (13mm, spanner) into the lower vibrating frame.
The lower vibrating frame is designated to move as the reciprocating motion.
4. Reinstall the Shaking table onto the unit in the reverse order.
Shaking Table
Shaking Table
- Remove the universal platform and miscellaneous
accessories from the Shaking table.
- Keep the working area enough for assembling.
- Follow the next instructions.
Upper Vibrating Frame
(Orbital Motion)
[Inside of Shaking System]
Make sure that a space (M5*43L) complied with our
standard dimension before using.
It may cause serious damage to the unit if it¡s too much
forced to install above parts.
Lower
Space
Install position
Lower Vibrating Frame
(Reciprocating Motion)
Page 27

Hide quick links:

Advertisement

Table of Contents
loading

This manual is also suitable for:

Lab companion sk-600Lab companion sk-71

Table of Contents