Jeio tech Lab Companion SK-300 Operating Manual page 39

Shaker
Table of Contents

Advertisement

Trouble related with main power
Trouble
Main power turns on
but
switch
lights
off
M ai n sw i t ch l i ght s
on but di spl ay does
not show
Whole
controlling
of the unit stops
without
cutting
main power or any
buttons
Pr ess
START/STO P
but t on but does not
w or k
Trouble related with mechanical shaking system
Trouble
Occur
trembleing
when
acceleration,
deceleration
and
uniform velocity
Taking
time
to
reach
the
desired
RPM
Make bumping noise
when operting
Differency
between
setting
RPM
and
speed
of
the
shaking table
Causes
Blackout
Main plug does not insert
correctly
Fuses are cut
Faulty connection of main
switch
Inferior main switch
Inner harness trouble
Poor transformer
Cotroller
and
trouble
Might
be
influenced
strong noise
Setting
values
controller trouble
Causes
The
shaker
is
equilibrated
Driving system troubles
Too many samples installed
on the shaking table.
Interference or loosing RPM
sensor
or
its
bracket
Interference
of
circuit lines
Driving system troubles.
Tension of the drive belt
is loosing
Solution
1. Check there might be blackout.
2. Main plug connects with socket correctly.
3. If fuses are cut, exchange new one.
4. If the problems do not solve, request
the service.
1. Request the service.
display
a
1. Avoid from the unit producing a strong
noise or put away that.
2. If the problems do not solve, request
the service.
of
the
1. After pressing SHAKE/SET button to input
RPM
setting
button to check operating condition.
2. If the problems do not solve, request
the service.
Solution
not
1. Set the unit to keep its balance,
refer to the chapter 2.
2. If the problems do not solve, request
the service.
1. Decrease the number of samples on the
shaking table.
1. Request the service.
fixing
inner
1. Request the service.
value,
press
START/STOP
Page 39