How to switch orbital motion to reciprocating motion
- Motion selectable, Factory default: Orbital motion

Switching motion orbital to reciprocal in SK-300

-
Upper vibrating frame
Lower vibrating frame
1. Remove the Shaking table from the unit.
Remove M5x10, wrench bolt (4mm, wrench handle) from the Shaking table.
2. Remove M5*23L, Spacer (13mm, spanner) from the upper vibrating frame.
The upper vibrating frame is designated to move as the orbital motion.
3. Install M5*43L, Spacer (13mm, spanner) into the lower vibrating frame.
The lower vibrating frame is designated to move as the reciprocating motion.
4. Reinstall the Shaking table onto the unit in the reverse order.
Page 26
Shaking Table
- Remove the universal platform and miscellaneous
accessories from the Shaking table.
Shaking table
- Keep the working area enough for assembling.
- Follow the next instructions.
Spacers for reciprocating motion
[Inside of Shaking System]
Make sure that a space (M5*43L) complied with our
standard dimension before using.
It may cause serious damage to the unit if it¡s too much
forced to install above parts.
Spacers for orbital motion
Spacers for
reciprocating motion