The Configuration Menu - Leica CV5030 Operating Manual

Robotic coverslipper
Table of Contents

Advertisement

5.4

The configuration menu

Press and
hold for 2 s
72
Press 1 x
Leica CV5030 – Robotic Coverslipper
The configuration menu is used to configure the parame-
ters with which the instrument works.
You can reach the configuration menu by pressing and
holding down the RESPOND button for two seconds.
The configuration menu consists of two submenus: MENU
A
(Fig.
71
The active submenu (71) is displayed in the right half of the
top display line.
In the left half of the upper display line, SET: indicates for
which parameter set (72) the current parameters are dis-
played.
There are four parameter sets (SET 1 through SET 4).
The bottom line in the display shows the following param-
eters in MENU A:
STK - Stroke length of the coverslipping mountant
VOL - Quantity of the coverslipping mountant
Fig. 30
TYP - Pressure in the dispensing unit
LEN - Correction of the stroke length (of the coverslipping
To toggle between the two submenus, press RESPOND.
In MENU B, you can select the following parameters:
CSP - Cover slip placement position
STP - Start position of the coverslipping mountant stroke
SPK - Volume of the audible warning signals and the
alarm
LIM - Adjustment of the limit value for cover slips
Fig. 31
30) and MENU B
(Fig.
mountant)
5. Operation
31).