4.3.4 Installing the dispensing group/dispenser

Dispensing group
37
40
35
39
Leica CV5030 – Robotic Coverslipper
36
35
38
Fig. 8
37
34
45
Fig. 9
4.
• Remove the dispensing group
packaging and keep the three nozzles
in a safe place.
• Insert the dispenser (35) into the holder (39)
for the prime position
• Connect the pressure hose (36) from the cov-
er of the mountant bottle (37) to the com-
pressed air output (41).
To do so, press the top white ring (42) down-
ward and insert the hose (36) as far as it will
go. Then pull the ring (42) back upwards. The
hose is then connected securely.
Enlarged
42
detail:
41
• Plug the cable (37) from the dispenser valve
into the socket (43) and screw it in place using
the knurled screw (38)
• Screw the cover (40) onto the mountant bottle
(44,
Fig.
10) and plug the mountant bottle into
the holder (45)
(Fig.
When screwing on the lid, make sure
that the sealing ring in the lid of the
mountant bottle fits correctly!