11.0 Notes on EMC

Where EMC is concerned, medical electrical equipment is subject to special safety measures and must be installed and
commissioned according to the EMC instructions stated herein.
ATTENTION:
The use of internal cables other than those specified in the Service Manual may result in increased emissions or decreased
immunity of the equipment.
ATTENTION:
The equipment should not be used adjacent to or stacked with equipment, other than with that which is intended for this
!
purpose. If adjacent or stacked use is necessary, the entire system should be observed to verify normal operation in the
configuration in which it will be used.
Guidance and manufacturer's declaration - electromagnetic immunity
The device is intended for use in the electromagnetic environment specified below. The customer or user of the device should
ensure that it is used in such an environment.
Emissions test
RF Emission acc.to
CISPR 11
RF Emission acc.to
CISPR 11
Harmonics acc. to
IEC 61000-3-2
Voltage
fluctuations/flicker
according to IEC
61000-3-3
Guidance and manufacturer's declaration - electromagnetic immunity
The device is intended for use in the electromagnetic environment specified below. The customer or user of the device should
ensure that it is used in such an environment.
Immunity test
Electrostatic discharge
(ESD) according to IEC
61000-4-2
Fast electrical transient/burst
IEC 61000-4-4
Compliance
Electromagnetic Environment - Guidance
Group 1
The device uses RF energy only for its internal function. Therefore, its RF
emissions are very low and it is unlikely that nearby electronic devices will be
affected.
Class B
The device is suitable for use in all establishments, including domestic and
those directly connected to the public low-voltage power supply network that
supplies buildings used for domestic purposes.
Class A
complies
IEC 60601- Test Level
± 6 kV Contact
± 8 kV Air
± 2 kV Mains
± 1 kV I/Os
Compliance Level
± 6 kV Contact
± 8 kV Air
± 2 kV Mains
± 1 kV I/Os
Electromagnetic Environment
- Guidance
Floors should be made of wood
or concrete or tiled with ceramic
tiles. If floors are non-conductive
synthetic, the relative humidity
should be at least 30 %.
Mains power quality should be
that of a typical commercial or
hospital environment.
33