2.0 For your safety

For your safety
• The ATMOS
S 41 Gyne is produced according to IEC
®
601 / EN 60601 and listed in the following classes:
- Class of protection 1
- Class IIa (EEC 93/42)
- Class IIb (installation of ATMOS
• The device may only be connected to a properly installed
safety socket.
• The ATMOS
S 41 Gyne is separated from the electricity
®
network via the rubber connector.
• Caution! Heated drawers may generate temperatures
above 40°C! Check the temperature of the instruments
prior to use. Contact service if necessary!
• The ATMOS
S 41 Gyne may only be used under the
®
supervision of skilled staff who have been authorised
by ATMOS and trained in its operation (IEC 601-1/EN
60601-1).
• The mains voltage indicated on the type plate must
correspond to the values of the supply network.
• Make sure prior to every application of the equipment that
it is technically safe and in proper condition. Damaged
leads and hoses must be replaced immediately!
• Please pay attention to the period tests in chapter 6.0
„Maintenance and service" on page 25.
• Correct configuration in assembly of country-specific
connections:
- green/yellow: protective conductor (PE)
- blue: neutral conductor (N)
- black or brown: phase (L)
• Please observe that the mobile parts could cause injury
and contusion!
• If there is a failure in the temperature control then higher
temperatures can occur. Check the temperature of the
instruments prior to use. Contact service if necessary!
6
!
RS 221!)
®
• Please note:
A medical insulating transformer with earth leakage
monitor or any similar safety system acc. to EN 60 601-1
is required, if several devices are connected over one
common power supply. The transformer must correspond
to the power consumption of all the devices to be
connected.
• The control panel must be clearly visible and accessible
for the user.
• The ambient conditions specified in the "Technical data"
(chapter 9.0) must be strictly observed!
• The suction system of the ATMOS
only be used for the suction of fluids in the medical field.
Never remove explosive, inflammable or corrosive gases
or fluids.
• Switch off all function modules when you finish your daily
business.
• The ATMOS
S 41 Gyne may be operated only in rooms
®
used for medical purposes, but not in areas subject to
explosion hazards and in oxygen rich environments.
• The ATMOS
S 41 Gyne meets the immunity to
®
interference requirements of IEC 601-1-2 / EN 60601-
1-2 „Electromagnetic Compatibility – Medical Electrical
Devices".
• The ATMOS
S 41 Gyne may not be operated with
®
devices not complying with the requirements of standard
EN 60601-1 „Medical Electrical Equipment" and EN
60601-1-2 „Electromagnetic Compatibility" (Medical
Electrical Equipment).
• There are no warranty claims whatsoever on defects
which arise from the use of third party accessories or
consumables.
• ATMOS is not liable for personal injury and damage to
property if
- no original ATMOS parts are being used,
- the advice for use in these operating instructions is
not being observed,
- assembly, new settings, alterations, extensions
and repairs have been carried out by personnel not
authorised by ATMOS.
• Also pay also attention to the safety information in
following chapters.
• The ATMOS
C 401 has been designed in accordance
®
with IEC 601/ EN 60601. It is assigned to VDE protection
class II. It may only be connected to a properly installed
earthed wall socket.