7.0 Troubleshooting

7.4 Suction system

Error indication
Weak suction or no suction at all
No suction, but vacuum gauge
indicates -0.7 bar
No suction and suction motor does
not start, control lamp is activated
If, nevertheless, the errors cannot be removed inform the ATMOS service staff. Do not start any attempts to repair the unit
yourself!
Also pay attention to corresponding chapters in respective operating instructions!
Possible cause
• Suction hose is clogged
• Auxiliary air control open
• Float of overflow safety closes the
suction opening
• Lid of secretion canister is not
closed tightly
• Bacterial filter is clogged
• Hose connections are leaking
• Connection hose is kinked
• Secretion penetrated the suction
pump
• Suction hose is clogged
• Float of overflow safety closes the
suction opening
• Bacterial filter is clogged
• Connection hose is kinked
• Pump / motor defective
Remedy
• Rinse the suction hose with water (hose
can also be removed)
• Check the auxiliary air control
• Check filling level in the secretion canister
• check that lid of the secretion canister is
closed properly
• Replace the bacterial filter
• Seal up or replace connections
• Check hose connections, remove kinks
• Have suction pump cleaned by a service
technician
• Rinse the suction hose with water (hose
can also be removed)
• Check filling level in the secretion canister
• Replace the bacterial filter
• Check hose connections, remove kinks
• Please contact your service technician
authorised by ATMOS