4.0 Operation
Fig. 7. SE 6501
4.3.2
ATMOS
SE 6501 (601.1900.0)
®
• Please pay attention to the safety and operating notes in
the attached operating instructions.
Whenever the electro-surgical device is used, the smoke
evacuation device automatically switches to extraction
operation. ATMOS
SE 6501 returns to basic suction
®
operation after completion of the incision procedure and the
tracking sequence.
Fig. 8.
4.3.2.1

Operating elements

Mains on/off switch
1
Mains filter
2
Value higher / value lower button to adjust the value in
3
the display
Indication of suction capacity
4
4.3.2.2
Switching on the ATMOS
Switch on the power switch ( fig. 8). The ATMOS
6501 performs a self-test. All the displays illuminate.
Suction performance and run-down time are adjustable.
Select the desired function.
Using the value higher / lower buttons you can change the
value of the selected function ( fig. 8).
You can see the value on the display. The displayed value is
either a percentage or in seconds. The percentage symbol
or seconds symbol illuminates. The percentage figure refers
to the device's maximum output (650 l/min).
!
!
SE 6501
®
SE
®
Automatic activation
With the integrated radiosurgical modules (ATMOS
221) and smoke gas module (ATMOS
is automatically activated/deactivated during application with
the radiosurgical unit.
Setting operation suction
The terms manual start and manual deactivation mean the
use of the <Start suction> button. Select the button <Value
higher> / <Value lower> a value between 20 and 100% and
confirm with the Start / Stop button. 100% corresponds to a
suction capacity of approx. 650 l/min.
Setting the follow-up time
The operation suction runs on for a specified length of time
after the ATMOS
SE 6501 is deactivated. This period of
®
time is the follow-up time. Select a value between 0 and 10
seconds, using the <Higher> / <Lower> ( fig. 8) buttons.
You can also let the operation suction continue running for
an indefinite period after deactivation of the electrosurgical
unit or after manual deactivation of the ATMOS
Select a period longer than 10 seconds, using the <Higher>
( fig. 8) button. The word "On" appears on the display. The
permanent operation suction mode is activated.
The permanent operation suction can be switched on and
off with the <Start suction> button ( fig. 8).
4.3.2.3

Suction

WARNING
Position the suction device, or the instrument with suction
device so that it cannot accidentally suck up swabs or the
like.
Press the <Start suction> button or activate the attached
laser or electro surgical device. The unit operates so long
as the button is pressed or the attached laser or electro
surgical device is activated. Suction then continues for the
duration of the set follow-up time, after which basic suction
takes place for the duration of the set basic suction time.
WARNING!
Never bring the suction hose into contact with the suction
site. It could become attached to the tissue.
ATTENTION:
Do not suck up any liquids. If there is a risk that you could
suck up liquids during the operation, attach an in-line filter
on the main filter. Replace the in-line filter immediately if you
have sucked up any liquid.
For smoke evacuation we recommend the use of the
ATMOS
specula.
®
RS
®
SE 6501) the suction
®
SE 6501.
®