Automated Thermal Cycler (ATC) System technical specifications: (continued)
Table 4
Feature
Dimensions
Weight
PCR volume range
Instrument memory
Power
[1]
Reaction volume @ 1 µL

System specifications

Table 5 Physical dimensions and weight, and power consumption
Dimension
Height
Width
Depth
Automated Thermal Cycler (ATC) System User Guide
Chapter 1 About the Automated Thermal Cycler (ATC) System
Height: 133.6 mm (5.3 in)
Width: 179.6 mm (7.1 in)
Depth: 317.5 mm (12.5 in)
Weight of Thermal Module: 6.0 kg (13.2 lb)
Weight of Control Module: 3.4 kg (7.48 lb)
Weight of 1m Interface Cable: 0.35 kg (0.77 lb)
100-240 V, 50-60 Hz Max: 600 W
Instrument footprint
Thermal module
Control box
133.6 mm (5.3 in) 76.4 mm (3.0 in)
179.6 mm (7.1 in) 257.0 mm
(10.1 in)
317.5 mm
342.0 mm
(12.5 in)
(13.5 in)

System specifications

Specification
Height: 76.4 mm (3.0 in) —
(including rubber feet)
Width: 257 mm (10.1 in)
Depth: 342.0 mm (13.5 in) —
(including air vents)
10−100 µL (96-well)
5−10 µL (384-well)
USB
Recommended clearance
Thermal module
N/A
N/A
Front: 51 mm
(2 in)
Rear: 102 mm
(4 in)
1
Control box
N/A
N/A
Front: 10 mm
(0.39 in)
Rear: 10 mm
(0.39 in)