A-2 Instrument Specifications
The following tables list the footprint and the weight of the instrument with the 96–Well
or 60–Well sample block module.
Footprint (With Sample Block Module Installed)
Height
26 cm (10 in)
Width
28 cm (11 in)
Depth
41 cm (16 in)
Note
You must provide sufficient space around the instrument for unrestricted air circulation.
Weight
Base Module
96-Well Gold/Aluminum Sample Block Modules
Note
See the sample block module user's manual for physical information on a module.
There is one version of the instrument. The power requirements of the instrument
under various power configurations are:
VAC ~100/120
8 AMP T (5x20 mm) or
8 AMP Slow Blow (3 AB)
VAC ~220/230/240
8 AMP T (5x20 mm)
IMPORTANT You must be able to disconnect the main power supply to the instrument
immediately if necessary.
In areas where the supplied power is subject to voltage fluctuations exceeding +/-10%
of the nominal value, a power line regulator may be required. High or low voltages can
have adverse effects on the electronic components of the instrument. The following
table specifies the electrical operating range for the instrument in various parts of the
world. Select appropriate fuse configuration based on the voltage used.
In Japan, the unit must have a dedicated 220-volt outlet! The unit will not
operate properly with a 100-volt outlet.
8.6 kg (19 lbs)
3.2 kg (7 lbs)
50/60 Hz
Use 250 V fuses
Max Power 725 VA