Operation
Erasing user-
defined programs
58
The individual data sections of the command line are separated by commas.
The dot "." has to be used as a decimal separator within a data section.
5
Enter the command line.
6
Finish the command line with Enter.
The data is transmitted to the pHotoFlex
If there is already a program stored under the selected number,
the programming is canceled. If you wish to store the program un-
der the selected number, you have to erase the program stored
under the number first.
After a successful transmission, the terminal program writes "!>".
After a failed transmission, the terminal program writes "?>".
To erase user-defined programs, enter a command line according to the fol-
lowing scheme in the terminal program:
Erase all user-defined
programs
Command line
U.520
Example
U.520
7
Enter the command line.
8
Finish the command line with Enter.
The data is transmitted to the pHotoFlex
The requested data is displayed on the terminal as the result.
After a successful transmission, the requested data is displayed on the ter-
minal.
After a failed transmission, the terminal program writes "?>".
pHotoFlex
®
STD.
Erase one user-de-
fined program
U.521#1,Program No
U.521#1,900
®
STD.