Zero adjust.
i Insert zero
cell (dist. water)
Cell
Start measurement
4
Insert the cell (see section 4.2).
Set another cell with < > < > and <START/ENTER> as neces-
5
sary.
6
Start the measurement of the zero adjustment with <START/EN-
TER>.
After measuring, the result of the zero adjustment is displayed and
stored.
i Zero adjust. successful! (successful zero adjustment) or i Calibra-
tion error! (zero adjustment not successful) is displayed as the result.
The zero adjustment is completed.
If i Calibration error! was displayed as the calibration result, a
note automatically reminds you of another zero adjustment be-
fore the next measurement.
Measuring is not possible without a valid zero adjustment.
7
Confirm the result with <START/ENTER>.
The zero adjustment is completed.
The meter is ready to measure.

4.5.8 Programs

You can view the most important data of all methods.
The method data is ordered according to the program number.
1
Open the menu Configuration/ Photometry / Programs.
The display shows the most important data of the selected program.
Programs
83: A6/25 MC
0.20 - 8.00 mg/l
NH4-N
16 mm
▲ ▼
i Scroll with
16 mm
Program number, test, code
Measuring range with unit
Citation form
Diameter of the cell
Operation