wtw pHotoFlex STD Manual page 53

Led filter photometer
Hide thumbs Also See for pHotoFlex STD:
Table of Contents

Advertisement

®
pHotoFlex
STD
Example,
output format CSV
ba75958e07
11/2020
The CSV output format delivers datasets separated by ";".
The data are output in the following order:
1
2
0;01.12.11;14:18:07;1;81;O2;[BV];[2 + 1];166;mg/l;VALID;;;
14;03.12.11;07:41:51;3;;;;;24.8;%;VALID;557;nm;VALID
1 Current number of the storage location
(or "0" during output from measured value display)
2 Date of storing
3 Time of storing
4 Adjusted ID
5 Program number (only for the measured parameter, Concentra-
tion)
6 Citation form (only for the measured parameter, Concentration)
7 BV: Measured parameter Concentration: Blank value
8 Dilution (only for the measured parameter, Concentration)
9 Main measured value or
Upper/lower measuring range limit
(only with measured value status, OFL/UFL)
10Unit of the main measured value
11Measured value status of the main measured value
* VALID: Measured value valid
* INVALID: Measured value invalid
* UFL: Measured value below the lower measuring range limit
* OFL: Measured value above the upper measuring range limit
12Secondary measured value or standard adjustment
for measured parameter Absorbance / % Transmission:
Wavelength of the measurement
for measured parameter Concentration: Status standard adjust-
ment [Cal]
13Unit of the secondary measured value
for the measured parameter, Absorbance% Transmission: nm
for the measured parameter, Concentration: Status standard
adjustment [Cal]
14Measured value status of the secondary measured value (see 11)
If the connected external printer does not print, please check
whether the same baud rate is set at the pHotoFlex
printer.
3
4 5 6
7
8
9
10
11
14
12
13
®
STD and
Operation
53