CHAPTER 3 Settings and Measurement
(2)-3 Ring capture setting
Ring Capture Function
Setting
(1) Ring/Relay capture
(2) Ring Capt. Pts.
(3) Ring Capt. Time
Captured data is saved into the File 1.
File1
When capturing is stopped at the STOP point in the above, the File4 and the File 5 are remained.
These files are consolidated into one file and it is saved. Then the ring catpuer is completed.
File4
Twice as many files as the Number of Ring Capture Points will be created at the maximum.
3-50
(1)
(2)
(3)
Set the capture function.
Off: The capture function is disabled.
Ring: Perform the ring recording. (For details, refer to the following figure.)
Relay: Data is captured consecutively for each file of up to 2000MB continuously.
To use the ring recording function, specify the number of data points per one file.
(For details, refer to the following figure.)
Display the measurement time that can be recorded in one file when the ring recording
function is enabled (On).
Start capturing data
When the specified number of data are saved into the File 1, the captured data is saved into the File 2.
The number of data is specified in the "Ring Capture Points" menu.
When the specified number of data are saved into the File 2,
the captured data is saved into the File 3.
File 1 is deleted at this point.
When the specified number of data are saved into the File 3,
the captured data is saved into the File 4.
File 2 is deleted at this point.
File2
File3
File4
File5
Description
When the specified number of data are saved into the File 4,
the captured data is saved into the File 5.
File 3 is deleted at this point.
* Repeat this action.
Stop capturing data
File5
Filexx••• : Saved