Gl840-Wv: B-565 - GRAPHTEC GL840 User Manual

Midi logger
Hide thumbs Also See for GL840:
Table of Contents

Advertisement

CHAPTER 4 Specification

GL840-WV: B-565

Item
Number of input channels
Input terminal type
Input method
Scan speed
Measurement ranges
Measurement accuracy
(23°C ±5°C)
• When 30 minutes or more have
elapsed after power was switched on
• Sampling 1 s/10 ch
• Filter ON (10)
• GND connected
4-6
20ch (200ch available when used with the extension terminal base)
* Possible to direct-connect or connect with the extension terminal connection cable
(sold separately) between the GL840 and terminal base, or between terminal
bases.
M3 screw type terminals (Rectangular flat washer)
Photo MOS relay scanning system
All channels isolated, balanced input
* Terminal b to be used to connect the Resistance bulb is shorted within all
channels.
10 ms/1 ch maximum
Voltage: 20, 50, 100, 200, 500 mV; 1, 2, 5, 10, 20, 50, 100 V; 1-5 V F.S.
Temperature
Thermocouples : K, J, E, T, R, S, B, N, C (WRe5-26)
Resistance bulb : Pt100, JPt100, Pt1000 (IEC751)
Temperature range: 100°C, 500°C, 2000°C
(In the case of Fahrenheit: 150°F, 750°F, 3000°F)
Humidity: 0 to 100% (voltage 0 V to 1 V scaling conversion) fixed
Voltage: ± (0.05% of F.S. + 10μV)
Temperature
Thermocouple
Measurement Temperature
Thermocouple
R/S
R : 300 < TS ≤ 1600℃
S : 300 < TS ≤ 1760℃
B
400 ≤ TS ≤ 600℃
600 < TS ≤ 1820℃
K
-200 ≤ TS ≤ -100℃
-100 < TS ≤ 1370℃
E
-200 ≤ TS ≤ -100℃
-100 < TS ≤ 800℃
T
-200 ≤ TS ≤ -100℃
-100 < TS ≤ 400℃
J
-200 ≤ TS ≤ -100℃
-100 < TS ≤ 100℃
N
-200 ≤ TS <
C (W)
Reference contact compensation accuracy
* Thermocouple diameters T - K: 0.32 φ, others: 0.65 φ
Resistance bulb
Measurement Temperature
Type
Pt100
JPt100
Pt1000
* 3-wire system
Description
Measurement Accuracy
Range (°C)
0 ≤ TS ≤ 100℃
±4.5℃
100 < TS ≤ 300℃
±3.0℃
±2.2℃
±2.2℃
±3.5℃
±2.5℃
±1.5℃
±0.8℃
±1.0℃
±0.8℃
±1.5℃
±0.6℃
±1.0℃
±0.8℃
100 < TS ≤ 1100℃
±0.6℃
0℃
±2.2℃
0 ≤ TS ≤ 1300℃
±1.0℃
0 ≤ TS ≤ 2000℃
±1.8℃
±0.3℃
Applied
Range (°C)
current
-200 ≤ TS ≤ 100℃
100 < TS ≤ 500℃
500 < TS ≤ 850℃
-200 ≤ TS ≤ 100℃
100 < TS ≤ 500℃
-200 ≤ TS ≤ 100℃
100 < TS ≤ 500℃
Measurement
1mA
±0.6℃
±0.8℃
±1.0℃
1mA
±0.6℃
±0.8℃
0.3mA
±0.6℃
±0.8℃

Hide quick links:

Advertisement

Table of Contents
loading

Table of Contents