CHAPTER 2 Checks and Preparation
2.21 Connecting the Illumination / Ultraviolet Sensor (GS-LXUV) (Optional)
For the specifications of illumination / ultraviolet sensor, refer to the Instruction Manual supplied with the
sensor.
The connecting procedure between the GL840 and illumination / ultraviolet sensor is described.
• Turn Off the GL840's power.
• Connect firmly the illumination / ultraviolet sensor's connector to the GL840's GS sensor and terminal /
adapter connection terminal (GS-INPUT).
• Turn On the GL840's power and make sure that the sensor settings are displayed on the AMP Setting
screen.
1. Hook portion ...................Used to mount to a wall.
2. Illumination sensor .........Sensor unit that measures illumination rays.
3. Ultraviolet sensor ...........Sensor unit that measures ultraviolet rays.
4. Connector .......................Used to connect to the connector on the GL840 module.
•
Please do not use in a state in which only the illumination / ultraviolet sensor is attached on a wall or the
like.
The damage of the connected cable and connector or the drop of the GL840 may be caused.
•
When measuring ultraviolet rays that have an effect on the eyes or the skin, wear protective glasses, apply
light shielding, etc.
•
If the sensor unit becomes dirty, this may affect measurements. Wipe it down with a soft cloth.
•
Make sure to handle the sensor unit in a way that it will not crack. If the sensor unit is cracked it may affect
measurements. Replace the sensor if cracked.
2-44
1. Hook portion
2. Illumination sensor
3. Ultraviolet sensor
4. Connector
GL840 main module