Operating Environment; Ambient Operating Conditions - GRAPHTEC GL840 User Manual

Midi logger
Hide thumbs Also See for GL840:
Table of Contents

Advertisement

Data Control & Processing
The application software provided lets you set conditions and monitor data on a PC.
The USB drive mode function enables the internal memory and SD CARD to be recognized as an external
drive by the PC.
(After connecting the GL840 to the computer, turn on the power of the GL840 while holding down the
[START] key.)
The WEB server function enables control and monitoring from a remote location without using dedicated
software
The FTP client function enables backup of measurement data to the FTP server.
The NTP client function enables synchronization of the time with the NTP server.
When optional wireless unit is installed, the data capture can be performed remotely by controlling the
GL100-WL.

1.3 Operating Environment

This section explains the operating environment for the GL840.

Ambient Operating Conditions

(1) Ambient temperature and humidity (the GL840 must be operated within the following ranges.)
• Temperature range: 0 to 45°C (0 to 40°C when a battery pack is mounted, 15 to 35°C when battery is
being charged)
• Humidity range: 5 to 85%R.H.
* When the screwless terminal (B-564SL) is connected to the GL840, it is necessary to use it in the
environment of the GL840.
(2) Environment (Do not use in the following locations.)
• Outdoor
• A Location such as being exposed to direct sunlight
• Locations exposed to salty air, corrosive gases, or organic solvents
• Dusty locations
• Locations subject to vibration or impact
• Locations subject to voltage surge or electromagnetic interference such as lightning or electric furnaces
(3) Installation category (over-voltage category)
• The GL840 belongs to Installation Category II defined in IEC60664-1.
• Never use the GL840 for Installation Category III or IV.
(4) Measurement category
• The standard terminal (B-564) and screwless terminal (B-564SL) connected to the GL840 should be not
used in Measurement Category II, III, and IV.
• The withstand high-voltage high-precision terminal (B-565) connected to belongs to Category II.
Therefore, it should be not use in Category III or IV.
(5) Altitude
• Altitude up to 2000 m.
(6) Mains supply voltage
• 100 to 240 VAC ±10%
CHAPTER 1 General Description
1-3

Hide quick links: