Maintenance
NOTICE
i
i
Since the use of certain cleaning and disinfecting
agents is not allowed among all countries, Dräger
tested and recommends several cleaning and
disinfecting agents. However, Dräger makes no
representations that such agents comply with or are
approved for use in your specific country's regulations.
Make sure to follow the procedure that is suitable for
your country.
For further information please contact Dräger or your
local dealer.
5.2.1
Cleaning and disinfecting procedure for the USA
only
1. Dismantling the device:
a. Separate breathing hose, respiratory inlet covering and
blower unit from each other.
b. Disconnect the carrying system from the blower unit.
c. If available, dismantle any accessories (e.g. hose and
device sleeves).
d. Dismantle splash guard cover and respiratory filter (see
section 5.3.3 on page 13).
2. Clean the respiratory inlet covering according to the
corresponding Instructions for Use.
3. Cleaning the breathing hose and carrying system:
a. Clean all parts with lukewarm water and 1008 GREEN
LIQUID HAND DISH WASH using a soft cloth.
b. Thoroughly rinse all parts under running water.
c. Prepare a disinfectant bath containing water and 800
SPUR-TEX Disinfectant (max. temperature: 86 °F/
30 °C, concentration: 2 fl.oz. per 1 gallon/60 ml per
3.79 l).
d. Place the parts to be disinfected into the disinfectant
bath (duration: 15 minutes).
e. Thoroughly rinse all parts under running water.
f. Dry all parts in the air or in a drying cabinet (max. tem-
perature: 140 °F/60 °C). Do not expose to direct sun-
light.
4. Clean and disinfect blower unit and splash guard cover
using disinfectant cloths (alcohol based).
In cases of strong contamination, the blower unit can be rinsed
under running water as follows:
1. Make sure the rechargeable battery remains inserted. Wa-
ter must not enter the battery compartment.
2. Close suction inlet and tube connection with protective
caps (available as accessories).
5.2.2
Cleaning and disinfecting procedure for other
countries
1. Dismantling the device:
a. Separate breathing hose, respiratory inlet covering and
blower unit from each other.
b. Disconnect the carrying system from the blower unit.
c. If available, dismantle any accessories (e.g. hose and
device sleeves).
d. Dismantle splash guard cover and respiratory filter (see
section 5.3.3 on page 13).
12
2. Clean the respiratory inlet covering according to the
corresponding Instructions for Use.
3. Cleaning the breathing hose and carrying system:
a. Prepare a cleaning solution containing water and NEU-
TRAL DISINFECTANT CLEANER (max. temperature
86 °F/30 °C, concentration: 1 fl. oz. per 2 gallons/30 ml
per 7.57 l).
b. Immerse the breathing hose and carrying system in the
solution (duration: 10 minutes). Make sure that all parts
are wetted.
c. Thoroughly rinse all parts under running water.
d. Dry all parts in the air or in a drying cabinet (max. tem-
perature: 140 °F /60 °C). Do not expose to direct sun-
light.
4. Clean and disinfect blower unit and splash guard cover
using disinfectant cloths.
In cases of strong contamination, the blower unit can be rinsed
under running water as follows:
1. Make sure the rechargeable battery remains inserted. Wa-
ter must not enter the battery compartment.
2. Close suction inlet and tube connection with protective
caps (available as accessories).
5.3

Maintenance work

5.3.1

Visual inspection

Check all parts thoroughly and replace damaged parts if
necessary. Check particularly the respiratory filter sealing
surface and O-rings of the blower unit for damage (e. g.
scratches) or contamination.
5.3.2

Replacing or charging the rechargeable battery

CAUTION
!
A short circuit could damage the rechargeable battery!
Make sure that the battery terminals do not come into
contact with metal during storage.
Removing the rechargeable battery:
1. Fold up carrying system if necessary.
2. Push battery lock button. Ensure that the rechargeable
battery does not fall down.
3. Remove rechargeable battery.
Inserting the rechargeable battery:
1. Fold up carrying system if necessary.
2. Position the two lower connection points of the
rechargeable battery at an angle in the battery
compartment sockets.
3. Fold the rechargeable battery in until it snaps audibly into
place.