3. Switch off the blower unit by pushing the button on the
control panel
for approx. 2 seconds.
4. Open the carrying system belt and take off the device.
5. Clean and disinfect the device (see section 5.2 on page 11).
4

Troubleshooting

4.1

Warnings

Fault
Cause
The residual run-
time of the
A segment of the
rechargeable
battery status
battery is low
indicator is flashing
(< 30 minutes).
yellow.
The particulate filter
residual capacity is
A segment of the
low (< 20 %).
particulate filter
residual capacity
indicator is flashing
yellow.
Malfunction during
switch-on (e. g.
caused by missing
A segment of the
hose or respiratory
flow rate indicator is
filter).
flashing yellow.
4.2

Alarms

4.2.1

Blower unit

Fault
Cause
The residual run-
time of the
A segment of the
rechargeable
battery status
battery is almost
indicator is flashing
exhausted
red.
(< 10 minutes).
The particulate filter
residual capacity is
A segment of the
almost exhausted
particulate filter
(< 10 %).
residual capacity
indicator is flashing
red.
Faulty breathing air
supply during
operation (e. g.
A segment of the
caused by missing
flow rate indicator is
hose, respiratory
flashing red.
filter or kinked hose).
General system
error
One segment at a
time is flashing red.
X-plore 8000
4.2.2
Fault
Status-LED flashing
red.
Remedy
Recharge the
battery soon or
replace with fully
5
charged battery (see
section 5.3.2 on
page 12).
5.1
Change particulate
or combination filter
soon (see section
5.3.3 on page 13).
Work to do
Verify device
Clean and disinfect the device
function and prepare
Visual inspection
again for use (see
section 3.2 on page
Replacing the O-ring at plug-in or bayonet-type
9).
hose connector
1
5.2
Remedy
Recharge the
battery or replace
with fully charged
battery (see section
5.3.2 on page 12).
Change particulate
or combination filter
(see section 5.3.3 on
page 13).
Verify device
function and prepare
again for use (see
section 3.2 on page
9).
Device must be
checked by Dräger
Service.

Standard battery charger

Cause
General error or
defect

Maintenance

Maintenance intervals

for gas-tight packed devices, otherwise every 6 months

Cleaning and disinfecting

CAUTION
!
Potential damage to components!
Do not use any solvents (e.g. acetone, alcohol) or
cleaning agents containing abrasive particles for
cleaning and disinfecting. Only use the methods and
the approved cleaning and disinfecting agents
described in this section. Other methods or agents can
cause damage to the components.
Health hazard!
Undiluted agents that come into direct contact with
eyes or skin are hazardous to health. Always wear
suitable protective equipment when working with the
cleaning and disinfecting agents described in this
section.
Troubleshooting
Remedy
Return rechargeable
battery to charger.
If the malfunction
occurs repeatedly
have Dräger Service
check battery and
charger.
1
X
1)
X
X