3.4
Dependency on air pressure
The anaesthetic partial pressure delivered by the Vapor 2000 (see
"Calibration" on page
5)
is virtually independent of the air pressure. Weather-related fluctuations are therefore
negligible. Altitude-related pressure changes between 700 and 1100 hPa, too, result in
minor deviations within the specified accuracy range (see technical specifications in
Instructions for Use). Consequently the physiological effect – the anaesthesia depth –
of a specific vaporizer setting is also independent of the air pressure.
When measuring the delivered concentration of the Vapor 2000 in partial pressure
(e.g. Dräger-IRIS or PM 8030/8035), no dependency on the ambient pressure occurs.
However, on measuring devices which give indications in percent by volume (e.g. Dräger-
PM 8020 or PM 8050) the measured values change with the air pressure. The measured
values increase when the air pressure falls below 1013 hPa.
Conversion formula:
Measured value [vol.%] . Air pressure [hPa]
Concentration [% partial pressure] = -------------------------------------------------------------------------------------------------
1013 hPa
Example:
At 4% partial pressure at an altitude of 1000 metres, 4.5 vol.% is displayed at 900 hPa (at
2000 metres 5.1 vol.% at 795 hPa).
Dependency of the boiling point of the anaesthetic on the air pressure (see
"Boiling point of
anaesthetic" on page
6).
Page 11
5327.400 Vapor 2000
05/99
Function Description