Fault
Operation:
Filling level not visible on
sightglass.
Anaesthetic in sightglass
discoloured.
5327.400 Vapor 2000
Cause
Vaporizer completely empty.
Vaporizer overfull.
Sightglass indicator
defective.
Halothane contains thymol,
which is enriched in the
vaporizer.
05/99
Remedy
Top up vaporizer.
Drain vaporizer down to
maximum mark. Check
concentration.
Repair (only by trained
specialists).
Drain off discoloured
anaesthetic and clean
vaporizer.
Fault-Cause-Remedy
Page 37